Laparoscopic
Closure of Trocar Defects
using Riza-Ribe® Needle
Product Link: Riza-Ribe®
Needle
Load ligature to Riza-Ribe® Needle by pushing plunger, threading
ligature in open loop and releasing plunger (Fig. B and C).
Fig. A Riza-Ribe® Needle

Fig. B |

Fig. C |
The Riza-Ribe®
Needles are used during laparoscopy for the Closure of Trocar Induced
Fascial Defects.
-
At
the end of laparoscopic procedure, select plug that matches your trocar
(e.g. 10 mm plug will match 10 mm trocar). Release trocar and remove
(Fig. 1).
- Immediately place
gloved finger over skin opening to retain pneumoperitoneum.
- Place R-Med®
Plug into defect (Fig. 2).
- Under Laparoscopic
visualization ensure that distal end of the R-Med® Plug is aligned
at the peritoneal surface (Fig. 3). This will align the Riza-Ribe®
Needle to pass 16-17 mm above the peritoneum, which will be above the
fascial level.
- Load Riza-Ribe®
Needle with a suture (e.g., Chromic, Vicryl O) as shown (Fig. B and
C).
- Thread the suture
loaded Riza-Ribe® Needle down the groove on one side of the R-Med®
Plug (Fig. 4) through fascia on the opposite side of the plug, into
the peritoneal cavity.
- Release the suture,
and withdraw the empty Riza-Ribe® Needle back through the R-Med®
Plug.
- Thread the empty
Riza-Ribe® Needle down the remaining groove, through fascia on the
opposite side of the plug, into the peritoneal cavity.
- Load the free suture
end into the Riza-Ribe® Needle (Fig. 5 and 6) with the help of a
grasper or a needle holder.
- Withdraw loaded
Riza-Ribe® Needle back through the R-Med® Plug, and release
the suture.
- After all sutures
are in place, remove the R-Med® Plug while tightening on the sutures
(Fig. 7).
- Lower the abdominal
pressure to reduce distention, allowing better approximation of the
fascia under lower stresses.
-
Tie sutures under direct visualization (Fig. 8).
-
Subcutaneous
and skin approximation can be performed to surgeon's choice.
-
For
Single Patient Use Only. Properly dispose of device in accordance
with local regulations.
CLOSURE OF MULTIPLE TROCAR DEFECTS
-
Place
another R-Med® Plug or matching plug into second defect (steps
1-3) to retain pneumoperitoneum.
-
Introduce
a 5mm grasper or needle holder through the axial bore of the first
R-Med® Plug.
-
Place
sutures in additional R-Med® Plugs as outlined above (steps 4-10).
|